www.skcltd.comFlite 2 High Volume Air Sampling PumpOperating InstructionsThis manual covers the following model:901-2011
Page 8 901-FTM Issue F www.skcltd.comGetting Started3) Use of the Keypad and LCD ScreenKey functions - - Switches between menu options and selects d
www.skcltd.com 901-FTM Issue F Page 9Getting StartedOn number entry menu screens such as the sample run time screen, each digit of the hour
Page 10 901-FTM Issue F www.skcltd.comProgramming the Flite 2 Pump1) Waking the Pump from Sleep ModeWhen not in use the pump should be put into ‘Sle
www.skcltd.com 901-FTM Issue F Page 11Programming the Flite 2 PumpSelect the ‘YES’ option in the ‘SET UP NOW ?’ menu screen, and press t
Page 12 901-FTM Issue F www.skcltd.comProgramming the Flite 2 PumpHaving set the required pump flow rate disconnect the flow calibrator from the sam
www.skcltd.com 901-FTM Issue F Page 13Programming the Flite 2 PumpTo enter a sample run time select the ‘YES’ option, and press to displ
Page 14 901-FTM Issue F www.skcltd.comProgramming the Flite 2 PumpPress to save the start delay time and display the next menu screen. Æ REPEAT RU
www.skcltd.com 901-FTM Issue F Page 15Programming the Flite 2 PumpTo check the pump flow rate select the ‘YES’ menu option and press . The pump wil
Page 16 901-FTM Issue F www.skcltd.comProgramming the Flite 2 PumpIf the sample is to be carried out at a later time, put the pump int
www.skcltd.com 901-FTM Issue F Page 17Operating the Flite 2 Pump1) Manual RunHaving programmed the desired flow rate and fitted a new filter into th
Page 2 901-FTM Issue F www.skcltd.comPurchase Details and Service HistoryThank you for choosing an SKC product. Your purchase is covered by our warr
Page 18 901-FTM Issue F www.skcltd.comOperating the Flite 2 PumpTo stop the pump at the end of the required sample period, press the and select
www.skcltd.com 901-FTM Issue F Page 19Operating the Flite 2 Pump3) Timed Run with Delayed StartHaving previously programmed a timed run with delayed
Page 20 901-FTM Issue F www.skcltd.comOperating the Flite 2 PumpAt the end of the first start delay time the pump will automatically sta
www.skcltd.com 901-FTM Issue F Page 21Operating the Flite 2 PumpTo restart the pump from Hold, press the keys in the sequence , , . The
Page 22 901-FTM Issue F www.skcltd.comTroubleshootingShould the Flite 2 pump detect that the battery is at a low charge state a low battery messag
www.skcltd.com 901-FTM Issue F Page 23Battery CareThe Flite 2 pump utilises sealed, valve regulated, lead/acid batteries. This type of
Page 24 901-FTM Issue F www.skcltd.comFlite 2 Pump ComponentsCasing foot1Casing screw2Casing screw2Casing lidBattery3Battery clamping bracket, nuts
www.skcltd.com 901-FTM Issue F Page 25Flite 2 Pump Head and Motor ComponentsPump head inlet tubeO ringValve housingValve diaphragmValve retainerTee
Page 26 901-FTM Issue F www.skcltd.comFlite 2 Pump Replacement PartsItem Part No. Description Item Part No. Description1 P901111 Casing feet (pack o
www.skcltd.com 901-FTM Issue F Page 27Flite 2 Pump AccessoriesPart No. Description901-210 Battery charger 100-240V ~ 50/60Hz - 12Vdc 1A with UK/EU/U
www.skcltd.com 901-FTM Issue F Page 1ContentsSpecifications ...
Page 28 901-FTM Issue F www.skcltd.comWarranty InformationLimited One Year Warranty1. SKC warrants that this instrument, and each of its c
www.skcltd.com 901-FTM Issue F Page 29Warranty Informationto the buyer. The remedy will consist of service or replacement of the instrument, or ref
SKC Limited 11 Sunrise Park, Higher Shaftesbury Road, Blandford Forum, Dorset DT11 8ST United KingdomT: +44 (0) 1258 480188 F: +44 (0) 1258 480184
Page 2 901-FTM Issue F www.skcltd.comSpecificationsDimensions ...178mm W x 115mm H x 206mm DWeight
www.skcltd.com 901-FTM Issue F Page 3SpecificationsPump FeaturesBattery capacity optionsChoose from two capacities of internal battery - 3.2
Page 4 901-FTM Issue F www.skcltd.comSpecificationsTypical Pump Run TimesFilter Type Flow RateRun Time P9012013.2Ah BatteryRun Time P9012027Ah Batte
www.skcltd.com 901-FTM Issue F Page 5General Information1) Care of the Flite 2 Pump• The Flite 2 pump is supplied complete with a rubber dust cap f
Page 6 901-FTM Issue F www.skcltd.comThe Flite 2 PumpInlet hosetail LCD screen Charging /power socketKeypadCarry handle Screw-in cap Casing feetCasi
www.skcltd.com 901-FTM Issue F Page 7Getting Started1) Fitting the BatteryThe Flite 2 pump is supplied without a battery. Refer to the instructions