Operating InstructionsSKC Inc.863 Valley View RoadEighty Four, PA 15330 USAForm 37740 Rev 1302

Seite 2 - Table of Contents

7Battery Operation2. Carefully align the ba ery jack on the replacement ba ery pack with the ba ery terminal on the bo om of the pump base plate

Seite 3 - AirChek 2000 Quick Guide

8Pump DisplayHOLD: Flashes when the pump is in HOLD mode (see page 10)SГ or S: Displayed when a sample time is manually programmed into pump memory

Seite 4 - Description

9Keypad BasicsThe AirChek 2000 Pump operates by pressing various bu on sequences on the keypad located on the front of the pump housing.  Scrolls

Seite 5 - Performance Profi le

10Pump Operating ModesRUN Pump is running; run time data is continuously updated in memory. The LCD shows real-time run time data. Press  to scroll t

Seite 6

11Verifying Battery Charge LevelThe LCD displays an icon that shows the current ba ery charge level (see Ba ery Operation on page 5). A new ba ery

Seite 7 - NiMH Battery

12ADJ Used during calibration with primary standard calibrator (not for use with CalChek feature). Press  to increase or  to decrease fl ow adjustme

Seite 8

13Time of day The hour will fl ash. Press  to increase and  to decrease hour. Press  to advance to minutes. Once minutes are fl ashing, press  to i

Seite 9

14Setting Sample Time (S Г)Program the AirChek 2000 from the keypad to run STEL, TWA, or any other run time from 1 to 999 minutes*. AirChek 2000 sampl

Seite 10 - Battery Operation

15Setting a Delayed StartThe Delayed Start feature is available in pump version 2.59 or higher.When se ing the pump manually for sampling to begin wi

Seite 11 - Introduction

16Deleting a Sample Time (SГ)To delete a sample time, enter User Interface Level Two and press the  bu on to scroll to SГ. Press  until SГ 00 appea

Seite 12

Description ...

Seite 13

17Flow Rate and Volume DisplayFlow rate displayed on the pump LCD is the fl ow to which the pump has been calibrated. To maintain fl ow as displayed, th

Seite 14 - Entering:

189. Press  until End appears on the LCD. Press [] to save fl ow se ing and exit the User Interface. The pump will remain in RUN. Reset run time d

Seite 15

19Calibration Using the CalChek Automatic Calibration FeatureThe CalChek automatic calibration feature is available when calibrating an AirChek 2000 P

Seite 16 - Setting Pump Flow Rate

20Calibration7. Place the pump in HOLD by pressing []. Press [] again to start the pump running, and enter the security code  in sequence on

Seite 17 - Setting Sample Time (S Г)

21CalibrationFull (Multiple-point) Calibration Using CalChekThis type of calibration provides fl ow correction across the basic operating range of the

Seite 18 - Deleting a Delayed Start

22CalibrationCalChek Full Calibration DataFull calibration data can be viewed and printed by going to the DataTrac 2000 Pump Manager window and clicki

Seite 19 - Deleting a Sample Time (SГ)

23CalibrationCalChek Error ChartSingle-point Calibration ErrorsError Problem TroubleshootingE41 Correction required too large. A mismatch of greater t

Seite 20 - Calibration

24Calibrating for Low Flow (5 to 500 ml/min) Sampling Applications Using a Constant Pressure Controller (CPC)For use with CPC and Adjustable Low Flow

Seite 21

25Sampling (1000 to 3250 ml/min Constant Flow)1. Following calibration, replace representative sampling medium with a new unexposed medium. Ensure r

Seite 22

26Low Flow Sampling (5 to 500 ml/min Constant Pressure)Low fl ow sampling requires a Constant Pressure Controller (CPC) and adjustable fl ow tube holder

Seite 23

Security Code Must be pressed within 10 seconds of previous commandMode ChangePress [] to toggle between Run and Hold.AirChek 2000 Quick GuideOp

Seite 24

27MaintenanceReplacing the Inlet Port Housing and Filter1. Remove the three screws that secure the inlet port housing to the top of the pump.2. Re

Seite 25

28This page is intentionally blank.

Seite 26

29Description Cat. No.Replacement Parts Ba ery Pack, NiCad P20136 Ba ery Pack, NiMH P20136MH Belt Clip P20139 Case P20137 Case with i

Seite 27

30Description Cat. No.Defender Primary Standard Calibrator, 50 to 5000 ml/min,includes ba ery, charger (100-240 V), so ware, and serial cable 717

Seite 28 - Sampling

31DataTrac 2000 SoftwareWith the optional DataTrac 2000 So ware accessory, the AirChek 2000 is programmable using a PC. DataTrac 2000 simplifi es chai

Seite 29

32SKC Limited Warranty and Return PolicySKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability an

Seite 30 - Maintenance

33IndexAC Charger ...5, 30Accessories ...30Activate Pump ...

Seite 31

1The SKC AirChek® 2000 Pump is an advanced programmable sample pump that combines lightweight compact design, computer-compatible circuitry, and a pat

Seite 32 - Replacement Parts

2Performance Profi leFlow Range: 1000 to 3250 ml/min (5 to 500 ml/min requires optional low fl ow adapter kit)Compensation Range: 3000 ml/min at 15

Seite 33 - Accessories

3Performance Profi leTime Display: Time of day in hours and minutes (12 or 24-hour clock) with AM and PM indicators Volume Display: Continually

Seite 34 - DataTrac 2000 Software

4Table 1. AirChek 2000 Run Time in Hours with NiCad BatteryFollowing are typical run times achieved when using a fully charged nickel-cadmium (NiCad)

Seite 35 - Warranty

5Installing the Battery Pack1. Carefully align the ba ery jack on the ba ery pack with the ba ery ter-minal on the bo om of the pump base plate a

Seite 36

6Battery Charge Level IndicatorThree bars indicate a full charge (normally appears a er charging),approximately 75 to 100%. Two bars indicate that th