Field Rotameters
This rotameter is packed to protect against damage during shipment. If damage
occurs, notify the carrier immediately. Also, check that you have received the
correct model and ow range required for your application.
Calibration
SKC rotameters are secondary standards and should be calibrated by the user
with a primary standard calibrator at regular intervals. Since rotameters are
a ected by temperature and atmospheric pressure, rotameter calibration should
be performed under the conditions of use or a mathematical correction should
be used.
Operation
Operate the rotameter in a vertical position to assure accuracy. Use it in a manner that
minimizes vibration and ow pulsation; both conditions may a ect the rotameter.
Flow rate is indicated by the point on the printed scale at which the center of the ball
oat comes to rest. The eld rotameter inlet must always be at atmospheric pressure
(i.e., exposed to the atmosphere). The rotameter is the last device connected to the
sampling train. Connect the inlet of the sampling pump to the outlet of the sampling
medium. Connect the inlet of the sampling medium to the outlet of the rotameter.
CAUTION
This rotameter is designed for use with non-hazardous gases at
pressures up to 100 psig and temperatures up to 65 C (150 F).
Do not use hazardous gases or liquids and do not exceed
temperature or pressure limits. Use with hazardous uids or
beyond pressure and temperature limits may cause failure,
which could result in personal injury.
Maintenance
Occasionally clean the rotameter if dirt appears in it or if oat movement is
restricted. To clean:
1. Remove the top plug, oat stop, and ball oat.
2. Wash the tapered hole, ball oat, oat stop, and top plug with a mild
liquid detergent and a so brush. Rinse all parts with clean water and
dry thoroughly. Avoid using solvents.
3. Reinstall the ball oat, oat stop, and top plug. Use a small amount of
halocarbon grease or any other compatible O-ring lubricant to ease
installation of the top plug and prolong O-ring life.
Speci cations
Float: Black glass or stainless steel ball
Body: Clear acrylic
Seals: Buna N O-ring
Pressure: 100 psig maximum
Temperature: 150 F (65 C) maximum
Fittings: Brass with hose connector, Tygon
®
tubing, and
Luer taper adapter
Form 37505 Rev 1208
* Full Scale
Product scale is marked in ccm (cc/min), which is equivalent to ml/min.
** Fittings adapted for use with HV30 Sample Pump
† Fittings adapted for use with QuickTake Sample Pumps
Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]
Meas. Range
(L/min)
Scale
(inch)
Accuracy*
±
Subdivisions
(L/min)
Cat. No.
0.05-0.5 2 5% 0.02 320-2A05
0.1-1.0 4 3% 20 320-4A1
0.4-5.0 4 3% 0.1 320-4A5
2-20 4 3% 0.5 320-4A20L
4-50 4 3% 1.0 320-440
3-30 4 3% 1.0 320-530**
3-30 4 3% 1.0 320-100†
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which
provides SKC’s sole liability and the buyers exclusive remedy. To view the com-
plete SKC Limited Warranty and Return Policy, go to h p://www.skcinc.com/
warranty.asp.
Seitenansicht 0
1

Inhaltsverzeichnis

Seite 1 - Operating Instructions

Field RotametersThis rotameter is packed to protect against damage during shipment. If damage occurs, notify the carrier immediately. Also, check tha

Kommentare zu diesen Handbüchern

Keine Kommentare