Operating Instructions
863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected]
IOM Personal Samplers and
IOM Samplers with MultiDust
Description
The patented* IOM Sampler is a sampling head that houses a reusable two-part
lter casse e with speci ed 25-mm lter for the collection of inhalable airborne
particles. The lter and casse e assembly are pre and post-weighed as a single unit
for gravimetric analysis. The IOM Sampler and casse e are available in conductive
plastic or stainless steel. The stainless steel model is ideal for sampling vapor-phase
isocyanates followed by chemical analysis. The IOM Sampler meets the following
U.S. and international standards:
ACGIH
®
sampling criteria for inhalable particulate mass
NIOSH Method 5700 for formaldehyde on dust
Australian standard for inhalable particulate
Preferred sampler for HSE Method MDHS 14/3 for inhalable and respirable dust
Complies with MDHS 25/3 for organic isocyanates (with stainless steel model)
Complies with MDHS 6 for lead (with accessory single-hole head)
• ISO
®
/CEN health-related fractions of bioaerosols
OSHA equivalent method for particulates not otherwise regulated (PNOR)
The sampler is supplied with a casse e cover and casse e transport clip to prevent
contamination during transport.
* U.S. Patent No. 4,675,034
Reference: OSHA le er November 8, 2011
IOM with MultiDust Foam Disc
The IOM can be used to sample for inhalable and respirable fractions simultane-
ously by using a lter and a MultiDust Foam Disc (polyurethane foam). In addition,
a polycarbonate lter can be combined with a MultiDust Foam Disc for e ective
size-selective bioaerosol sampling and be er microorganism survivability when
compared to lter-only sampling.
1
Conductive plastic IOM
Stainless steel IOM
Seitenansicht 0
1 2 3 4 5 6 ... 11 12

Inhaltsverzeichnis

Seite 1 - IOM Samplers with MultiDust

Operating Instructions 863 Valley View Road, Eighty Four, PA 15330 USATel: 724-941-9701 Fax: 724-941-1369 e-mail: [email protected] Personal S

Seite 2 - Performance Profi le

10Post-weighing for Inhalable and Respirable FractionWear powder-free gloves when handling casse es and use forceps when working with fi lters.1. Equ

Seite 3 - Preparation and Assembly

116. Create fi ve 10-fold serial dilutions (to 10-5 of original suspension) from each suspension and spread 0.1 ml of each dilution (plus 0.2 ml of und

Seite 4 - General Weighing Guidelines

12www.skcinc.comForm 37372 Rev 1302Ordering InformationDescription Cat. No.IOM Sampler and cassette,§ in conductive plastic, with transport clip and c

Seite 5 - Sampling

2Collar clipO-ring CFrontplateCassette topFilterCassette bottomHousing bodyO-ring BO-ring AFigure 1. Confi guration for conductive plastic IOM Sampler

Seite 6 - Transporting Loaded Cassettes

3Preparation and AssemblyHandlingWear powder-free gloves when handling casse es and use forceps when working with fi lters and foams to prevent the tr

Seite 7 - Calibration

4IOM Stainless Steel Casse eUse stainless steel casse e only.Autoclave stainless steel casse e up to 392 F (200 C).Reminder: Use powder-free gloves

Seite 8 - Sampling for Bioaerosols

5Filter Cassette: Opening, Installing Filter, and Closing (Figure 4)1. Separate the casse e top from the casse e bo om.a. Plastic casse e: Use g

Seite 9 - Analysis

65. Place a single MultiDust Foam Disc for respirable sampling (Cat. No. 225-772) into the inlet of the casse e top so that it sits just below the r

Seite 10 - and Analysis

7CalibrationCalibrate pump fl ow rate using a primary standard calibrator, IOM Calibration Adapter accessory, and a representative loaded IOM Sampler i

Seite 11 - References

85. A er sampling, stop the pump. Unscrew the front plate from the housing body and remove the casse e from the sampler (may use sterile forceps and

Seite 12 - Ordering Information

9Transporting Samples to a LaboratoryWear powder-free gloves when handling casse es and use forceps when working with fi lters. Users may wish to use